Products

PICO-RHEED

Low Emission Reflection High-Energy Electron Diffraction

Product Information

The PICO-RHEED allows crystal structure analysis of organic thin films because of enabling use with low-emission current.

PICO-RHEED can significantly reduce sample damage caused by electron beams due to low -emission current, while enabling observation of high-brightness reflection electron diffraction images by using a dedicated MCP screen.

By replacing the MCP screen with a standard (optional), PICO-RHEED can be used as a standard RHEED to analyze inorganic thin films.

Features
  • Significantly reduced sample damage (1/500 to 1/2800: in-house comparison)
  • Variable bias for Wehnelt electrode
  • High reliability due to low noise and high stability output
  • Compatible with kSA400 RHEED image/video analysis system
Drawings

Specifications
Electron Gun
Model No.RDA-005G
Insulation VoltageDC 30 kV
Beam Spot Dia.Φ90 μm (min.)
FilamentTungsten hairpin
Wehnelt ElectrodeVariable bias
Focusing LensCoreless magnetic lens
Deflection LensToroidal electromagnetic lens
Beam DeflectionX/Y: ±8°
Magnetic ShieldStandard feature
Working Pressure< 1 × 10-3 Pa
Differential Pumping type <1 Pa*
Max. Bake Temperature200 ℃
Mounting FlangeICF70-FH
SizeΦ100 mm × 335 mm
(455 mm including connector)

MCP Screen for PICO-RHEED
Model NumberRDA-100M
Effective Screen DiameterΦ77 mm
Viewport7056 Glass
(Radiation shielding cover included)
Mounting FlangeICF152-FH
Screen ShutterManual rotation type
MCP Power Supply for PICO-RHEED
Model NumberRDA-101P
MCP Output Voltage(Ch1)1 kV (max.)
Fluorescent Screen
Output Voltage(Ch 2)
   4 kV (max.) 
Power Supply VoltageAC 100 V ~ 240 V  

Power Supply for PICO-RHEED
Model NumberRDA-006P

※Specifications and dimensions are subject to change without notice.

PAGETOP